The CA202 piezoelectric accelerometer is one of the core products of the vibro-meter brand under Meggitt, designed specifically for industrial vibration monitoring and measurement. The sensor employs a symmetrical shear-mode polycrystalline piezoelectric sensing element housed in an austenitic stainless steel case, offering excellent mechanical stability and long-term reliability. Model 144-202-000-205 is the standard version equipped with a 3-meter integral low-noise cable, suitable for industrial vibration monitoring in non-hazardous areas.


With its high sensitivity, wide frequency response, strong environmental adaptability, and long-term stability, the CA202 sensor is widely used in vibration monitoring and fault diagnosis systems in industries such as power generation, petrochemical, aerospace, and heavy machinery. Its design complies with multiple international standards and can achieve high-precision vibration signal acquisition and analysis when paired with compatible signal conditioners (e.g., IPC70x series).



2. Key Features & Benefits


Feature Description
High Sensitivity 100 pC/g, suitable for high-precision measurement of weak vibration signals.
Wide Frequency Response 0.5 Hz to 6,000 Hz, covering the vibration frequency range of most industrial equipment.
Wide Operating Temperature Range The sensor can operate continuously from -55°C to +260°C, suitable for extreme temperature conditions.
Robust Sealed Construction Hermetically welded austenitic stainless steel case and flexible stainless steel protection hose, resistant to dust, water, oil, and corrosion.
Symmetrical Differential Output Internal case insulation with differential signal output for strong noise immunity.
Integral Cable Low-noise cable is hermetically welded to the sensor, eliminating connection point failures.
Suitable for Harsh Environments Optional explosion-proof (Ex) versions for use in potentially explosive atmospheres (Zone 0/1/2).
Easy Installation Standard M6 mounting holes, no additional insulating washers required, compatible with various mounting adapters.




3. Application Areas

The CA202 accelerometer is suitable for the following industrial vibration monitoring scenarios:

  1. Power Generation Equipment

    • Vibration monitoring of bearings and casings in steam turbines, gas turbines, wind turbines, and hydro generators.

  2. Oil & Gas

    • Vibration analysis and fault prediction for compressors, pumps, fans, and pipelines.

  3. Heavy Machinery & Manufacturing

    • Vibration monitoring of machine tools, stamping equipment, conveyor systems, and metallurgical equipment.

  4. Aerospace & Defense

    • Vibration testing on engine test benches, ground support equipment, and propulsion systems.

  5. Transportation

    • Vibration diagnosis for rail vehicles and marine propulsion systems.

  6. Research & Laboratories

    • Structural dynamics testing, modal analysis, and vibration table control.



4. Ordering Information

The CA202 is available in multiple versions and cable lengths. Below are some example models:

Version Type Cable Length Part Number (PNR)
Explosion-Proof (Ex) 3m 144-202-000-106
Explosion-Proof (Ex) 6m 144-202-000-116
Standard Version 3m 144-202-000-205
Standard Version 6m 144-202-000-216
Standard Version 11m 144-202-000-226
Standard Version 20m 144-202-000-236

For other cable lengths or special configurations, please contact your local Meggitt representative.



5. Installation & Usage Recommendations

  1. Mounting Orientation: The sensor should be mounted along the measurement axis, with the arrow direction aligned with the vibration direction to ensure a positive signal output.

  2. Mounting Surface: Should be flat and clean; no additional insulating washer is required.

  3. Cable Securing: Avoid excessive bending or strain on the cable; use cable clamps for support.

  4. Signal Conditioning: Must be used with a charge amplifier (e.g., IPC70x) to obtain a voltage signal for data acquisition systems.

  5. Calibration: Dynamically calibrated at the factory (5 g, 120 Hz); generally no recalibration is needed.


6. Recommended Accessories

Accessory Name Model Part Number (PNR) Purpose
Mounting Adapter Kit MA133 809-133-000-011 For CA20x/CE31x series, includes Micaver (mica-glass) thermally isolating base


Table 1: General Specifications

Item Specification
Product Line vibro-meter CA202
Example Model 144-202-000-205 (Standard Version, 3m cable)
Sensing Principle Shear-mode piezoelectric effect
Output Type Charge output (requires charge converter)
Signal Connection 2-core shielded cable, insulated from case
Compatible Signal Conditioner IPC70x series charge amplifier

Table 2: Electrical & Dynamic Performance (@23°C)

Parameter Value Remarks
Sensitivity 100 pC/g ±5% Calibrated at 120 Hz, 5 g
Dynamic Measurement Range 0.01 – 400 g (peak)
Overload Capacity (peak) ≤500 g
Linearity Error 0.01–20 g: ±1%
20–400 g: ±2%

Transverse Sensitivity ≤3%
Resonant Frequency >22 kHz (nominal)
Frequency Response Range 0.5 – 6,000 Hz ±5% Lower limit depends on signal conditioner
Insulation Resistance ≥10⁹ Ω
Sensor Capacitance (pin-to-pin) 5,000 pF
Cable Capacitance (per meter) 105 pF (pin-to-pin)
210 pF (pin-to-ground)

Table 3: Environmental Specifications

Parameter Sensor Integral Cable
Continuous Operating Temperature -55°C to +260°C -55°C to +200°C
Short-Term Survival Temperature -70°C to +280°C -62°C to +250°C
Temperature Sensitivity Error -55°C ~ +23°C: 0.25%/°C
+23°C ~ +260°C: 0.1%/°C
Relative to 23°C reference
Protection Rating Hermetically welded, resistant to dust, water, oil, salt spray
Case Material Austenitic stainless steel (1.4441)
Protection Hose Material Heat-resistant stainless steel (1.4541)
Base Strain Sensitivity 0.15 × 10⁻³ g/µε (@250 µε peak-to-peak)
Shock Resistance ≤1000 g (half sine, 1 ms)

Table 4: Mechanical & Mounting Specifications

Parameter Value
Case Material Austenitic stainless steel
Weight (sensor) Approx. 250 g
Weight (cable) Approx. 135 g/m
Mounting Method 4×M6×35 Allen screws with spring lock washers
Mounting Torque 15 N·m (11.1 lb-ft)
Cable Length (Model 205) 3 meters
Cable Type MTLN-205 (K205A) low-noise shielded cable
Wire Colors Blue (signal+), White (signal-), Green (case ground)

Table 5: Certifications & Compliance

Certification Type Standard/Certificate
CE Marking Complies with EU Directives
EAC Marking Eurasian Customs Union Conformity Certification
Electromagnetic Compatibility EN 61000-6-2:2005, EN 61000-6-4:2007+A1:2011
Electrical Safety EN 61010-1:2010
Environmental Compliance RoHS (2011/65/EU)
Explosion-Proof Certification (Optional) ATEX, IECEx, UKEX, cCSAus, etc. (see Ex versions for details)


Previous: 
Next: 

Quick Links

PRODUCTS

OEM

Contact Us

 Telephone: +86-181-0690-6650
 WhatsApp: +8618106906650
 Email:  sales2@exstar-automation.com / lily@htechplc.com
 Address: Room 1904, Building B, Diamond Coast, No. 96 Lujiang Road, Siming District, Xiamen Fujian, China
Copyright © 2025 Exstar Automation Services Co., Ltd. All Rights Reserved.